Changing the Fuse Configuration, if necessary
The Well-plate Handler is delivered for 230–240V operation using two 1.0 Amp
fuses. To convert the unit from the dual 1.0 Amp fuse configuration used for
230-240V operation to the single 1.5 Amp fuse configuration used for 100-120V
operation, refer to the diagrams and instructions below. Otherwise, proceed to
"Seating the Well-plate Handler on its Locating Plate"
Electrical shock hazard.
WA RN ING
Disconnect the power cord before changing the fuses.
For continued fire protection and correct functioning of the unit, replace fuses only
with fuses of the same type and rating.
1 Ensure the Well-plate Handler's power switch is in the OFF position and the
2 Using a small flat-blade screwdriver, gently pry the cover/fuse block
Figure 14
3 Grasp the voltage selector card (not the indicator pin which is attached to
18
power cord is unplugged.
assembly away from the power entry port housing and remove (see
Figure
14).
Prying open the cover/fuse-block assembly
it) using needle-nosed pliers and remove it from the housing (see
Figure
15).
on page 21.