5
Troubleshooting and Error Handling
A plate must be installed in the transport position of the Automation Interface.
NO TE
Eject Plate
A plate must be installed in the well-plate autosampler.
NO TE
Open Door
Close Door
Test Barcode Reader
The Reset button resets the Automation Interface to its initial state.
Testing with the Hand-Held Controller
The Automation Interface Tests screen (see
selection of commands that allow you to test or adjust the Automation
Interface.
Figure 56
80
Opens the door and transfers the plate from the
sampler into the Automation Interface.
Opens the door.
Closes the door.
Reads the barcode on the plate on the Automation
Interface transport and displays the result in the field.
Automation Interface Tests screen
Figure
56) gives access to a
1100 Series Sample Capacity Extension