Checks for Sequence Table columns:
Column Name
An MS Excel file for sequence import may look similar to
1100 Series Sample Capacity Extension
Location
Sample Type
Method
Inj/Location
CalLevel
UpdateRF
UpdateRT
Interval
Data Filename
Dilution, Multiplier, ISTD
Amount, Sample Amount
Inj Volume
Study Name (only if
ChemStore is installed)
Working with the Sample Capacity Extension
Valid Range
Must be a valid location string, e.g. Vial 1, 2, P1-A1, P1-B-2, etc.
The validity of the Location with the configured plates is not
checked.
Sample, Calibration or Control Sample; the check is not
case-sensitive.
It is not necessary that the method exists, although this check is
made. Inclusion of the extension .m is optional; other extensions
are ignored.
1 – 99 or blank.
1 – 99 or blank.
No Update, Average, Replace, Bracket or Delta%; the check is not
case-sensitive.
No Update, Average or Replace; the check is not case-sensitive.
1 – 999 or blank.
Inclusion of the extension .d is optional; other extensions are
ignored.
Any positive value or blank.
0 – Max. Inj. Vol or blank
The study must exist in order for custom fields to be imported.
4
Figure 47
69