Modifying the Well-plate Handler for 10-inch Racks
If you intend to use the Well-plate Handler with 10-inch racks instead of
8.5-inch racks, you need to make the following modifications:
1 Unscrew the button extension from the underside of the gripper arm (see
Figure 20
You can move the gripper arm while the Well-plate Handler is switched off without
NO TE
damaging the instrument.
2 Remove the two Phillips-head screws from the gripper fingers at each side
3 Install the new gripper finger set that was delivered with the 10 inch rack
Store the button extension and the gripper fingers for the 8.5-inch racks carefully in a
CAU TI O N
safe place, in case you need to convert the Well-plate handler back to 8.5-inch rack
configuration.
1100 Series Sample Capacity Extension
Figure
20).
The button extension (left) and removed (right)
of the gripper, and remove the fingers.
kit, with the fingers pointing down. Ensure that all screws are tight.