Agilent Technologies 1100 Series Reference Manual page 48

Sample capacity extension
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2
Installation
The teaching commands are separated into two groups: Rotary Teaching
calibrates the rotary positions, Vertical Teaching calibrates the vertical
positions. In addition, the Reset Well-plate Handler button moves the gripper
arm to its rotary and vertical home positions. The Initialize Counter button is
used to reset the WPH Movements counter in the Module Details of the
Well-plate Sampler to zero, for example after maintenance or exchange of the
Well-plate Handler.
The rotary home position is not directly above position R1.
NO TE
Rotary Teaching
The Rotary Teaching group contains a list of the coordinates for each of the rack
positions and the Automation Interface transport position. You select a
position to calibrate, for example, R1, and use the buttons to set the
coordinates. The Well-plate Handler arm can be rotated about 330 degrees, via
a stepper motor with 3400 steps. Positive numbers move the arm in a
clockwise direction; negative numbers move the arm counter-clockwise.
Plus 10
Plus
Minus
Minus 10
Move Home Vertical
Move Down
Set Default Values
Move Plate Test
40
Moves the Well-plate Handler arm 10 steps clockwise.
Moves the Well-plate Handler arm one step clockwise.
Moves the Well-plate Handler arm one step
counter-clockwise.
Moves the Well-plate Handler arm 10 steps
counter-clockwise.
Moves the Well-plate Handler arm to its topmost
position.
Moves the Well-plate Handler arm 1.5 cm downwards.
Sets the factory-default coordinates.
Moves the Well-plate Handler arm down until it
encounters a plate, picks up the plate, moves the arm to
the vertical home position, moves the arm down,
releases the plate and moves the arm back to its vertical
home position.
1100 Series Sample Capacity Extension