Commissioning A New Sensor - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Measured variable: chlorine
Table of Contents

Advertisement

Operating menus
Sensor status
[OK] = - 500 nA ... + 500 nA
[WRN] Warning = - 500 to - 1000 nA[ LOW ZERO] or + 500 to + 1000
nA[ HIGH ZERO]
[ERR] error: ≦ - 1001 nA or ≧ + 1001 nA
5.
Press the
[CAL=ACCEPT] : The calibrated value for the sensor is transferred to the con‐
ð
troller. The continuous display appears again.
[OK=RESET] : The controller is reset to factory-determined zero point. This is
useful, for example, when connecting a new sensor.
[ESC] : The calibration process is cancelled. The controller will continue to
operate with the old values. The continuous display appears again.
6.
[OK=RESET] : Press the
Only with
ð The calibrated value for the sensor is transferred to the controller. The contin‐
uous display appears again.
8.1.3

Commissioning a new sensor

1.
Carry out the
process and
brate sensor zero point [CHECK
ZERO]' on page 51 in doing so, use
[OK=RESET] to set the zero point
[0] or calibrate the zero point with
to
[CAL=ACCEPT] .
2.
Now calibrate the sensor slope.
Ä Chapter 8.1.1 'Sensor slope cali‐
bration' on page 48
ð The menu
[OK=RESET] : Press the
3.
Only for
key
ð The calibrated value for the
sensor is transferred to the
controller. The continuous dis‐
play appears again.
52
key
[CHECK ZERO]
Ä Chapter 8.1.2 'Cali‐
[INPUT] appears.
key