Error Messages - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Measured variable: chlorine
Table of Contents

Advertisement

Maintenance
10.2

Error messages

User qualification for diagnosis: trained user, see
n
tions' on page 10 . Further qualifications depend on the type and scope of possible
troubleshooting measures to be carried out.
Error display delay
Various errors are only displayed after approx. 10 seconds after switching on the
controller.
Tab. 7: Error messages
Display
Description /
cause
RANGE ↓
Sensor current
insufficient
RANGE↑
Sensor current
excessive
T RANGE
Measured tem‐
perature beneath
measuring range
T RANGE
Measured tem‐
perature above
measuring range
CAL
No valid user cal‐
ERROR
ibration exists
CHECK‐
Control checkout
TIME
time elapsed
mA
mA output cur‐
RANGE ↑
rent has an
upper limit
mA
mA output cur‐
RANGE ↓
rent has a lower
limit
78
Ä Chapter 2.2 'Users' qualifica‐
Status
Mode
1
2
Basic load
Error
Basic load
Error
Basic load
Error
Basic load
Error
-
Error
Basic load
Error
-
Error
-
Error
Measured
Correction
variable
variable
output
3
output
Error cur‐
-
rent
Error cur‐
-
rent
Error cur‐
Error current
rent
Error cur‐
Error current
rent
-
-
Error cur‐
-
rent
-
-
-
-
4