Calibrate:summary:output:current:level; Calibrate:summary:output:current[:Data]; Calibrate:save - ITech IT6500C Series Manual

Highpowered and programmable power supply
Hide thumbs Also See for IT6500C Series:
Table of Contents

Advertisement

Command syntax
CALibrate:EQUal:OUTPut:CURRent[:DATA] <NRf>
Arguments
NRf
Unit
A
Example
CAL:EQU:OUTP:CURR 12.0

CALibrate:SUMMary:OUTPut:CURRent:LEVel

This command is used to set calibration point of summary current.
Command syntax
:
CAL:SUMM:OUTPut:CURRent:LEVel <P0|P1>
Arguments
P0|P1
Example
CAL:SUMM:OUTP:CURR:LEV P0

CALibrate:SUMMary:OUTPut:CURRent[:DATA]

This command is used to set calibration data of summary current.
Command syntax
CALibrate:SUMMary:OUTPut:CURRent[:DATA] <NRf>
Arguments
NRf
Unit
A
Example
CAL:SUMM:OUTP:CURR 20.0

CALibrate:SAVe

This command is used to save the calibration parameters in a nonvolatile
memory.
Copyright© Itech Electronics Co., Ltd.
Calibration Command
102

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the IT6500C Series and is the answer not in the manual?

Questions and answers

Table of Contents