4.3.5 IR Voltage Calibration (IRV)
Keep the setting on the Valhalla voltmeter at DC and 2kV range. IR calibration is applicable to
the Sentry 30 Plus instrument.
Press [F1] = UP to go to
Press [STOP] to return instrument to steady state.
Press [START] to get offset value.
Press [F1] = INC or [F2] = DEC until the Sentry Plus display reads the same as the voltmeter.
Press [ENTER] to accept reading.
Sentry Plus reverts to "IRV Offset" display (incorporating new cal value).
Press [F1] = UP to go to
Press [STOP] to return instrument to steady state.
Press [START] to get full value.
Press [F1] = INC or [F2] = DEC until the Sentry Plus display reads the same as the voltmeter.
Press [ENTER] to accept reading.
Sentry Plus reverts to "IRV Full" display (incorporating new cal value).
Service & Calibration
Cal Step 5: 'IRV 1kV Offset
Cal Step 6: 'IRV 1kV Full
(50V)'.
(1kV)'. Change voltmeter range to 20kV.