Programming a GC Test (continued):
CURR:
Set
HIGH:
DWELL
DWELL:
Time
0.1 - 1s
CURR:
Set
HIGH:
LOW
DWELL:
Resistance Limit
0 - 5Ω
CURR:
HIGH:
[F1]
:
Program Step 2
DWELL:
or
[F4] : Exit Program Mode
1 - 10
END GC Test Programming. After selecting the LOW resistance limit, one can either press
[F1] = NEW to change the step number in the highlighted box (1-10) and start programming
STEP 2-10 OR one can press [F4] = EXIT to exit programming function and return to STAND
BY status.
For this example, press [F1] = NEW and proceed to program step 2 as an AC Hipot Test (¶ 2.5).
Page 36 of 85
STEP 1/1
GC
0.1A
1.2Ω
0.3s
RMT
LOCK
STEP 1/1
GC
0.1A
1.2Ω
0.5s
RMT
LOCK
STEP:
1/1
GC
0.1A
1.2Ω
0.5s
RMT
LOCK
LOW
:
OFF
INC.
DEC.
NEXT
EXIT
OFST
ERR
LOW
:
OFF
INC.
DEC.
NEXT
EXIT
OFST
ERR
LOW
:
0.3Ω
NEW
MORE..
NEXT
EXIT
OFST
ERR
To set Dwell time:
F1
0.1 - 1s in 0.1s increments
F2
F3
To move highlighted box to LOW.
F4
To set Low resistance limit:
F1
0 - 5Ω in 0.1Ω increments.
F2
F3
To move highlighted box to STEP.
F4
F1
To go to Step 2.
F2
F3
To
exit
programming mode and return
F4
to STAND BY status.