Programming A Multi-Step Test - QuadTech Sentry Plus Series Instruction Manual

Hipot testers
Table of Contents

Advertisement

2.10

Programming a Multi-Step Test

Paragraphs 2.4, 2.5, 2.6, 2.7 and 2.8 illustrate the programming of a 5-step test (GC, AC, DC, IR
& PA).
To program a multiple step test, power-up the Sentry Plus Series instrument so the
'Stand By' display is shown (Figure 2-2).
parameters for the first test (GC). After specifying LOW resistance limit, press [F3] = NEXT and
[F1] = NEW to enter test parameters for the second test (AC). At this time, do NOT press [F4] =
EXIT because this will exit you from programming mode. After specifying FALL time for the
second test press [F3] = NEXT to continue entering test parameters for the third test (DC).
Continue this process for up to 10 steps. When finished entering desired number of test steps,
press [F4] = EXIT to exit programming mode and return to Stand By status.
In Stand By status, to view the test steps just programmed press [F4] = MORE. Figure 2-4
illustrates the Stand By display and resultant test steps display if the programming example in
paragraphs 2.4-2.8 was followed.
STAND BY
with
Memory ID
Continuing with the 5-step example, press [F4] = MORE to view
the 5 test steps (GC, AC, DC, IR & PA) stored to location 1.
1
2
Test
3
Steps
4
5
Operation
M01
1/5
GC
0.1A
1.2Ω
0.5s
STAND BY
RMT
OUTPUT
MEASURE
M01
GC
0.1A
1.2Ω
AC
2.750kV
15.00mA
DC
2.500kV
2.999mA
IR
0.750kV
1.5MΩ
PA
OFF
STAND BY
RMT
Figure 2-4: View Programmed Test Setups
Press [F1] = PROGRAM and enter the test
LOW
:
0.3Ω
PROGRAM
PRESET
MENU
MORE..
LOCK
OFST
ERR
RESULT
------------
------------
------------
------------
------------
LOCK
OFST
ERR
F1
F2
F3
F4
F1
F2
OFFSET
F3
EXIT
F4
Page 47 of 85