Programming an AC Hipot Test (continued)
VOLT
Set
HIGH
Arc Limit
TIME
1 - 20mA 0 = OFF
VOLT
Set
HIGH
Ramp Time
TIME
0 - 999s 0 = OFF
VOLT
Set
HIGH
Fall Time
TIME
0 - 999s 0 = OFF
Set
Next
VOLT
Test Step
HIGH
TIME
[F1]
:
Program Step 3
or
[F4] : Exit Program Mode
1 - 10
END AC Hipot Test Programming. After selecting the FALL Time, one can either press [F1]
= NEW to change the step number in the highlighted box (1-10) and start programming STEP 3-
10, OR one can press [F4] = EXIT to exit programming function and return to STAND BY
status.
For this example, press [F1] = NEW and proceed to program step 3 as a DC Hipot Test (¶ 2.6).
Page 38 of 85
STEP 2/2
AC
:
2.750kV
:
15.00mA
:
10.0s
RMT
STEP 2/2
AC
:
2.750kV
:
15.00mA
:
10.0s
RMT
STEP 2/2
AC
:
2.750kV
:
15.00mA
:
10.0s
RMT
STEP
2/2
AC
:
2.750kV
:
15.00mA
:
10.0s
RMT
LOW
:
3.00mA
ARC
:
OFF
RAMP
:
OFF
FALL
:
OFF
LOCK
OFST
ERR
LOW
:
3.00mA
ARC
:
OFF
RAMP
:
OFF
FALL
:
OFF
LOCK
OFST
ERR
LOW
:
3.00mA
ARC
:
OFF
RAMP
:
3.0s
FALL
:
OFF
LOCK
OFST
ERR
LOW
:
3.00mA
ARC
:
OFF
RAMP
:
3.0s
FALL
:
3.0s
LOCK
OFST
ERR
Set arc current limit:
INC.
F1
1-20mA in 0.1mA increments
DEC.
F2
NEXT
F3
To move highlighted box to RAMP.
EXIT
F4
To set Ramp Time:
INC.
F1
0 - 999s in 0.1s increments
DEC.
F2
NEXT
F3
To move highlighted box to FALL.
EXIT
F4
To set Fall Time:
INC.
F1
0 - 999s in 0.1s increments
DEC.
F2
NEXT
F3
To move highlighted box to STEP.
EXIT
F4
NEW
F1
To set next test step = 3
MORE..
F2
NEXT
F3
To
exit
programming mode and return
EXIT
F4
to STAND BY status.