2.5
Programming an AC Hipot Test
This test is applicable to the Sentry 10, 20 and 30 Plus instruments. With the instrument in
'stand-by' status, press [F1] = PROGRAM. If continuing the 5-step example from ¶2.4, you are
already in program mode and on the AC test mode page. Follow the
right side of this diagram to program the individual AC hipot test parameters.
STEP 2/2
VOLT
:
[F1] = New
HIGH
:
Program
TIME
:
Step 2 = AC
Voltage is 0
STEP 2/2
VOLT
:
Select
HIGH
:
AC
TIME
:
Test Mode
Select Mode
STEP 2/2
VOLT
:
Set
HIGH
:
Test Voltage
TIME
:
0.05 - 5kV
STEP 2/2
VOLT
:
Set
HIGH
:
High Current
TIME
:
Limit
0.001 - 20mA
STEP 2/2
Set
VOLT
:
Test Time
HIGH
:
TIME
:
Test time = 0
then continuous
voltage until
STOP
0, 0.1 - 999s
is pressed.
STEP 2/2
Set
VOLT
:
Low Current
HIGH
:
Limit
TIME
:
0 - 20mA 0 = OFF
Continue on next page.
Operation
AC
LOW
ARC
0.000kV
RAMP
0.500mA
FALL
3.0s
RMT
LOCK
AC
LOW
ARC
0.000kV
RAMP
0.500mA
FALL
3.0s
RMT
LOCK
AC
LOW
ARC
0.000kV
RAMP
0.500mA
FALL
3.0s
RMT
LOCK
AC
LOW
ARC
2.750kV
RAMP
0.500mA
FALL
3.0s
RMT
LOCK
AC
LOW
ARC
2.750kV
RAMP
15.00mA
FALL
3.0s
RMT
LOCK
AC
LOW
ARC
2.750kV
RAMP
15.00mA
FALL
10.0s
RMT
LOCK
:
OFF
FIRST
:
OFF
:
OFF
MORE..
:
OFF
NEXT
EXIT
OFST
ERR
:
OFF
UP
:
OFF
:
OFF
DOWN
:
OFF
NEXT
EXIT
OFST
ERR
:
OFF
INC.
:
OFF
:
OFF
DEC.
:
OFF
NEXT
EXIT
OFST
ERR
:
OFF
INC.
:
OFF
:
OFF
DEC.
:
OFF
NEXT
EXIT
OFST
ERR
:
OFF
INC.
:
OFF
:
OFF
DEC.
:
OFF
ENTER
EXIT
OFST
ERR
:
OFF
INC.
:
OFF
:
OFF
DEC.
:
OFF
NEXT
EXIT
OFST
ERR
green
arrows (!) on the
F1
To change value in highlighted box.
F2
To delete or insert a test step.
To move highlighted box around
F3
display to select parameter to change
F4
To exit programming mode.
F1
Set MODE = AC
F2
F3
To move highlighted box to VOLT
F4
To set test voltage:
F1
0.05 - 5kV in .001kV increments
F2
F3
To move highlighted box to HIGH.
F4
To set high current limit:
F1
0.001 - 20mA in .001mA increments
F2
F3
To move highlighted box to TIME.
F4
To set test time:
F1
0, 0.1 - 999s in 0.1s increments
F2
F3
To move highlighted box to LOW.
F4
To set low current limit:
F1
0 - 20mA in 0.01mA increments
F2
F3
To move highlighted box to ARC.
F4
Page 37 of 85