Remote - QuadTech Sentry Plus Series Instruction Manual

Hipot testers
Table of Contents

Advertisement

3.1

Remote

A 9-pin D-Series remote control connector is located on the rear panel of the Sentry Plus Series
instrument. There is a black 5 screw terminal strip for the remote input signals: START,
RESET, COM and INTERLOCK. Inputs require a contact closure. Figure 3-1 illustrates the
Remote terminal strip connector and 9-pin D-Series connector.
Before connecting the instrument to its power source, the interlock function on the rear panel
remote connector (terminal strip) must be properly utilized. This is an important safety feature
for the protection of the operator. When the INTERLOCK jumper is removed, there is no high
voltage at the OUTPUT. Therefore, to initiate a test make sure the interlock jumper is in place.
UNDER
TEST
1
6
PASS
9-pin D-Series Remote Connector
The Sentry Plus Series instrument has three output signals on the rear panel. The UNDER TEST
relay is closed during a test. The PASS relay is closed when the DUT is judged GOOD. The
FAIL relay is closed when the DUT is judged NO GOOD. These relays are rated for voltage up
to 115VAC and current <100mA.
Interface
Section 3: Interface
RESET
START
5
9
FAIL
5-screw Remote Terminal Strip
Figure 3-1: Sentry Plus Series Remote Connectors
START
RESET
COM
INTER
LOCK
Page 67 of 85

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sentry 10 plusSentry 20 plusSentry 30 plus

Table of Contents