Error Messages (FAIL result)
When the measurement value was judged NO GOOD and FAIL is shown on the display, an error
message denoting the test result will be shown on the display also. Table 2-4 lists the possible
error messages for a NO GOOD/FAIL judgment.
Error Message
HIGH FAIL
LOW FAIL
ARC FAIL
I/O FAIL
VOLTAGE OVER
CURRENT OVER
GFI TRIPPED
NO OUTPUT
INRUSH FAIL
MEMORY FULL
Operation
Table 2-4: Error Messages
Description
Measured value (current or resistance) is over the programmed high limit.
Measured value (current or resistance) is below the programmed low limit.
Current arc is over the programmed high limit.
Hardware Current Limit exceeded. Check DUT for short.
Voltage reading is greater than the hardware allowable # of digits.
Current reading is greater than the hardware allowable # of digits.
Ground Fault Interrupt is No Good.
Output is too Low. (Try increasing test time)
Charging Current is over limit. (Check test fixture)
Greater than 60 tests/slot have been stored.