Setup Menu; Voltage - QuadTech 1865 Instruction Manual

Megohmmeter/ir tester
Table of Contents

Advertisement

2.6.2 Setup Menus
Setup
The first of the three main menus is Setup, shown in Figure 2-9. Each function controls a
measurement condition and is described in detail below.

2.6.2.1 Voltage

Setup
Accepts entry (up to 6 digits and decimal) of a test voltage between 1 and
1000 volts. This is the voltage applied to the unit under test during the test cycle and
present at the DUT connection anytime that the CAUTION HIGH VOLTAGE lamp is
ON. As in the case of many of the menu entries on the 1865 instrument, an "error
message" will be displayed for an invalid entry.
Operation
I/O
Utilities
Voltage
= 1000
Charge Time
= 0
Dwell Time
= 0
Measure Time
= 0
Discharge Time
= 0
Mode
Auto Manual
Range
> >
Limit
= 100 e+ 9
Stop on Pass
= 0
#
to Average
= 0
Figure 2-9: Setup Menu
I/O
Utilities
1000
Voltage
=
Charge Time
= 10
Dwell Time
= 10
Measure Time
= 10
Discharge Time
= 10
Mode
Auto Manual
Range
> >
Limit
= 100 e+ 9
Stop on Pass
=
0
#
to Average
=
0
Figure 2-10: Test Voltage Entry
e
Page 37 of 87