The In-Rush current limit for the DC hipot test is a low limit which is monitored during initial charging of the device
under test. When activated (other than Off), current below the programmed value is considered to be a Fail
condition. This current monitoring is generally used to check that the charging current to the device exceeds a
specified value (> programmed limit), thus ensuring that the device is properly connected to the tester.
2.7
Programming an IR (Insulation Resistance) Test
This test is applicable to the Sentry 30 Plus instrument. With the instrument in 'stand-by' status,
press [F1] = PROGRAM. If continuing the 5-step example from ¶2.6, you are already in
program mode and on the AC test mode page. This example shows a GC test as Step1, an AC
test as Step 2, a DC test as Step 3 and how to program an IR test in Step 4. Follow the
arrows (!) on the right side of this diagram to program the individual IR test parameters.
STEP
4/4
VOLT
:
0.000kV
[F1] = NEW
HIGH
:
0.500mA
Program
TIME
:
3.0s
Step 4 = IR
Voltage is 0
STEP 4/4
VOLT
:
0.000kV
Select
HIGH
:
0.500mA
IR
TIME
:
3.0s
Test Mode
Select Mode
STEP 4/4
VOLT
:
0.000kV
Set
LOW
:
1.0MΩ
Test Voltage
TIME
:
3.0s
0.05 - 1kV
STEP 4/4
Set
VOLT
:
0.750kV
Low
LOW
:
1.0MΩ
Resistance
TIME
:
3.0s
Limit
0.1MΩ - 50GΩ
STEP 4/4
VOLT
:
0.750kV
Set
LOW
:
1.5MΩ
Test Time
TIME
:
3.0s
0, 0.1 - 999s
Continue on next page.
Operation
NOTE: In-Rush Current
AC
LOW
:
OFF
ARC
:
OFF
RAMP
:
OFF
FALL
:
OFF
RMT
LOCK
OFST
AC
LOW
:
OFF
ARC
:
OFF
RAMP
:
OFF
FALL
:
OFF
RMT
LOCK
OFST
IR
HIGH
:
OFF
RAMP
:
OFF
DWELL
:
OFF
FALL
:
OFF
RMT
LOCK
OFST
IR
HIGH
:
OFF
RAMP
:
OFF
DWELL
:
OFF
FALL
:
OFF
RMT
LOCK
OFST
IR
HIGH
:
OFF
RAMP
:
OFF
DWELL
:
OFF
FALL
:
OFF
RMT
LOCK
OFST
FIRST
F1
MORE..
F2
NEXT
F3
EXIT
F4
ERR
UP
F1
DOWN
F2
NEXT
F3
EXIT
F4
ERR
INC.
F1
DEC.
F2
NEXT
F3
EXIT
F4
ERR
INC.
F1
DEC.
F2
NEXT
F3
EXIT
F4
ERR
INC.
F1
DEC.
F2
NEXT
F3
EXIT
F4
ERR
green
To change value of highlighted box.
To delete or insert a test step.
To move highlighted box to MODE.
To exit programming mode.
Set MODE = IR
To move highlighted box to VOLT
To set test voltage:
0.05 - 1kV in 0.001kV increments
To move highlighted box to LOW.
To set low resistance limit:
0.1M - 50GΩ
in 0.1MΩ increments
To move highlighted box to TIME.
To set dwell time:
0, 0.1 -999s in 0.1s increments
To move highlighted box to HIGH.
Page 41 of 85

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sentry 10 plusSentry 20 plusSentry 30 plus

Table of Contents