G16 International Power Strip - QuadTech Sentry Plus Series Instruction Manual

Hipot testers
Table of Contents

Advertisement

3.2

G16 International Power Strip

The Sentry Plus Series instruments can be connected to the G16 International Power Strip as
illustrated in Figure 3-6 for safety testing of many European corded products.
*Australia
*North America
*Sweden
*Austria
1. Connect the three G-16 ground connectors to the Sentry Plus RTN/LOW terminal.
2. Note: there is a second RTN/LOW terminal on the rear of the Sentry instrument.
3. Connect the white banana plug to the Sentry Plus OUTPUT terminal.
4. Plug DUT's AC power cord into corresponding module on G16 power strip.
STOP
START
1
0
QuadTech
AUSTRALIA
UNITED KINGDOM
Figure 3-6: G16 International Power Strip Connection to Sentry 30 Plus
Page 70 of 85
*United Kingdom
*Norway
*Germany
*Switzerland
Sentry 30 Plus Hipot Tester
Q
uadTech
Sentry 30 Plus AC/DC/IR Hipot Tester
CAL
UPDATE
PASS
FAIL
G16 International Power Strip
DENMARK
NORTH AMERICA
*Denmark
*Finland
*Netherlands
*Italy
F1
F2
F3
F4
DANGER
RTN/LOW
!
CAUTION
Max 5kVAC
6kVDC
Black
Green/
Banana
Yellow
Lug
to
RTN
or RTN
on Rear
Panel
NORWAY
FINLAND
SWEDEN
GERMANY
AUSTRIA
NETHERLANDS
F1
F2
F3
F4
OUTPUT
White
Custom
Plug
Banana
Plug
to
to
RTN
HV
OUTPUT
Green
Lug
to
RTN
SWITZERLAND
ITALY
Interface

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sentry 10 plusSentry 20 plusSentry 30 plus

Table of Contents