Programming an DC Hipot Test (Continued)
STEP 3/3
VOLT
Set
HIGH
Arc Limit
TIME
1 - 5mA 0 = OFF
STEP 3/3
VOLT
Set
HIGH
Ramp Time
TIME
0 - 999s 0 = OFF
STEP 3/3
VOLT
Set
HIGH
Dwell Time
TIME
0 - 999s 0 = OFF
STEP 3/3
VOLT
Set
HIGH
Fall Time
TIME
0 - 999s 0 = OFF
STEP 3/3
VOLT
Set
HIGH
I-Rus Limit
TIME
0, 0.5uA-5mA 0 = OFF
Set
Next
VOLT
Test Step
HIGH
TIME
[F1]
:
Program Step 4
or
[F4] : Exit Program Mode
1 - 10
END DC Hipot Test Programming. After selecting the I-RUSH limit, press [F1] = NEW to change the step
number in the highlighted box (1-10) and start programming STEP 4-10, OR press [F4] = EXIT to exit
programming function and return to STAND BY status.
For this example, press [F1] = NEW and proceed to program step 4 as an IR Test (¶ 2.7).
NOTE:
The In-Rush current range is dependent on the programmed High Limit. Refer to Safety Features in the
Page 40 of 85
DC
:
2.500kV
:
2.999mA
:
5.0s
RMT
DC
:
2.500kV
:
2.999mA
:
5.0s
RMT
DC
:
2.500kV
:
2.999mA
:
5.0s
RMT
DC
:
2.500kV
:
2.999mA
:
5.0s
RMT
DC
:
2.500kV
:
2.999mA
:
5.0s
RMT
STEP
3/3
DC
:
2.500kV
:
2.999mA
:
5.0s
RMT
Specifications section, page 8.
LOW
:
0.015mA
ARC
:
OFF
RAMP
:
OFF
DWELL
:
OFF
FALL
:
OFF
I-RUS
:
OFF
LOCK
OFST
ERR
LOW
:
0.015mA
ARC
:
OFF
RAMP
:
OFF
DWELL
:
OFF
FALL
:
OFF
I-RUS
:
OFF
LOCK
OFST
ERR
LOW
:
0.015mA
ARC
:
OFF
RAMP
:
3.0s
DWELL
:
OFF
FALL
:
OFF
I-RUS
:
OFF
LOCK
OFST
ERR
LOW
:
0.015mA
ARC
:
OFF
RAMP
:
3.0s
DWELL
:
3.0s
FALL
:
OFF
I-RUS
:
OFF
LOCK
OFST
ERR
LOW
:
0.015mA
ARC
:
OFF
RAMP
:
3.0s
DWELL
:
3.0s
FALL
:
3.0s
I-RUS
:
OFF
LOCK
OFST
ERR
LOW
:
0.015mA
ARC
:
OFF
RAMP
:
3.0s
DWELL
:
3.0s
FALL
:
3.0s
I-RUS
:
OFF
LOCK
OFST
ERR
Set arc current limit:
INC.
F1
1-5mA in 0.1mA increments
DEC.
F2
NEXT
F3
To move highlighted box to RAMP.
EXIT
F4
To set Ramp Time:
INC.
F1
0 - 999s in 0.1s increments
DEC.
F2
NEXT
F3
To move highlighted box to DWELL.
EXIT
F4
To set Dwell Time:
INC.
F1
0 - 999s in 0.1s increments
DEC.
F2
NEXT
F3
To move highlighted box to FALL.
EXIT
F4
To set Fall Time:
INC.
F1
0 - 999s in 0.1s increments
DEC.
F2
NEXT
F3
To move highlighted box to I-RUS.
EXIT
F4
Set in-rush current limit:
INC.
F1
0, 0.5uA -5mA
in 0.001mA increments
DEC.
F2
NEXT
F3
To move highlighted box to STEP.
EXIT
F4
NEW
F1
To set next test step = 4
MORE..
F2
NEXT
F3
To
exit
programming mode and return
EXIT
F4
to STAND BY status.