2.4

Programming a Ground Continuity Test

This test is applicable to the Sentry 10, 20 and 30 Plus instruments. A Ground Continuity (GC)
test is usually done first to verify the ground connection before high voltage is applied in the AC
& DC Hipot tests. With the instrument in 'stand-by' status, press [F1] = PROGRAM. The AC
test mode programming screen is displayed. Follow the
this diagram to program the individual GC test parameters.
VOLT:
0.000 kV
In STAND BY:
HIGH:
0.500 mA
Press [F1] =
TIME:
PROGRAM
Voltage is 0
CURR:
Select
HIGH:
GC
DWELL:
Mode
Select Mode
CURR:
Set
HIGH:
Current
DWELL:
0.1A
CURR:
Set
HIGH:
HIGH
DWELL:
Resistance Limit
0.1 - 5Ω
Continue on Next Page.
Operation
STEP:
1/1
AC
3.0 s
RMT
LOCK
STEP 1/1
GC
0.0 A
1.0 Ω
0.3 s
RMT
LOCK
STEP 1/1
GC
0.0A
1.0Ω
0.3s
RMT
LOCK
STEP 1/1
GC
0.1A
1.0Ω
0.3s
RMT
LOCK
green
arrows (!) on the right side of
LOW
:
OFF
ARC
:
OFF
RAMP
:
OFF
MORE..
FALL
:
OFF
NEXT
EXIT
OFST
ERR
LOW
:
OFF
UP
DOWN
NEXT
EXIT
OFST
ERR
LOW
:
OFF
INC.
DEC.
NEXT
EXIT
OFST
ERR
LOW
:
OFF
INC.
DEC.
NEXT
EXIT
OFST
ERR
F1
F2
F3
To move highlighted box to AC.
F4
F1
To select Mode = GC
F2
F3
To move highlighted box to CURR.
F4
F1
To set current to 0.1A
F2
F3
To move highlighted box to HIGH.
F4
To set High resistance limit:
F1
0.1 - 5Ω in 0.1Ω increments.
F2
F3
To move highlighted box to DWELL.
F4
Page 35 of 85

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sentry 10 plusSentry 20 plusSentry 30 plus

Table of Contents