Specifications - QuadTech Sentry Plus Series Instruction Manual

Hipot testers
Table of Contents

Advertisement

Dielectric Strength
Sentry 10, 20, & 30 Plus
AC Output Voltage:
Voltage Display:
AC Current Display:
Sentry 20 & 30 Plus
DC Output Voltage:
Voltage Display:
DC Current Display:
Insulation Resistance
Sentry 30 Plus
Insulation Resistance:

Specifications

Range:
0.05 to 5kV AC, in 1V steps
± (1% of setting +5V)
Regulation:
Frequency:
50/60Hz selectable
± (1% of reading +5V)
Accuracy:
Resolution:
1Volt
Range:
0.001mA to 20mA AC, in 1µA steps
± (1.5% of reading + 5 counts) (Total)
Accuracy:
Range:
0.05 to 6kV DC, in 1V steps
± (1% of reading +5V)
Accuracy:
Resolution:
1Volt
Range:
0.0001mA to 5mA DC
Resolution:
0.1µA
± (1.5% of reading + 5 counts)
Accuracy:
Voltage:
50 - 1000V DC in 1V steps
± (1% of reading + 5V)
Accuracy:
Range:
0.1MΩ - 50GΩ (voltage dependent)
0.1MΩ - 1GΩ, ± (10% + 5counts) < 100V
Accuracy:
0.1MΩ - 1GΩ, ± (7% + 5counts) < 500V
1MΩ - 1GΩ, ± (4% + 5counts) ≥ 500V
1GΩ - 10GΩ, ± (7% + 5counts) ≥ 500V
10GΩ - 50GΩ, ± (12% + 5counts) ≥ 500V
Page 7 of 85

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sentry 10 plusSentry 20 plusSentry 30 plus

Table of Contents