QuadTech Sentry Plus Series Instruction Manual page 20

Hipot testers
Table of Contents

Advertisement

Condensed Operating Instructions (Continued)
Connection to Device under Test (DUT)
Figure COI-4 illustrates the connection of the Sentry Plus Series unit to a single DUT using the
S02 1-meter HV cable set that comes standard with the instrument. The custom white banana
plug/red alligator clip is connected between the OUTPUT terminal on the Sentry Plus Series unit
and the high side of the device under test. The black banana plug/alligator clip is connected
between the RTN/LOW terminal on the Sentry Plus Series unit to the low side of the DUT.
STOP
START
1
0
Measurement Mode
1. Turn [POWER] ON.
2. Allow Sentry Plus Series instrument a 15-minute warm up time.
3. Connect S02 Black ground cable to Sentry Plus Series unit RTN/LOW terminal
4. Connect S02 White/red HV cable to Sentry Plus Series unit OUTPUT terminal.
5. Press [F1] = PROGRAM and enter test parameters. When finished programming, press
[F4] = EXIT to return to STAND BY status.
6. Press [F4] = MORE to access Offset function. Press [F3] = OFFSET. Follow Offset
instructions. When Offset is complete, press [F4] = MORE to return to STAND BY.
7. Connect device under test (DUT) to test leads.
8. Press [START].
9. Record measurement.
10. Press [STOP].
Page 20 of 85
Q
uadTech
Sentry 30 Plus AC/DC/IR Hipot Tester
M1
STEP 1/5
AC
1.250kV
0.500mA
3.0s
STAND BY
RMT
LOCK
CAL
UPDATE
PASS
FAIL
COI-4: Connection to Device under Test
LOW: OFF
PROGRAM
ARC: OFF
RAMP: OFF
PRESET
FALL: OFF
MENU
MORE..
OFST
ERR
DANGER
!
CAUTION
Max 5kVAC
6kVDC
F1
F1
F2
F2
F3
F3
F4
F4
OUTPUT
RTN/LOW
S02
Cable
Set
-
+
DUT