Dsp Assembly Replacement; Tools Required; Removal Procedure; Replacement Procedure - Agilent Technologies E5061B Service Manual

Network analyzers
Hide thumbs Also See for E5061B:
Table of Contents

Advertisement

Replacement Procedure

DSP Assembly Replacement

DSP Assembly Replacement

Tools Required

Removal Procedure

Refer to Figure 4-9 for this procedure.
Step 1. Remove the Outer Cover as described in "Outer Cover Replacement" on page 54.
Step 2. Remove the Front Panel as described in "Front Panel Replacement" on page 56.
Step 3. Disconnect the flat cable (item 1) from the DSP Assembly.
Step 4. Disconnect the two S-ATA cable (item 2) from the DSP Assembly.
Step 5. Remove the two 5/8 inch nuts and the two washers (item 3) from the BNC connectors.
Step 6. Remove the two TORX T10 screws (item 4) fastening the DSP Assembly.
Step 7. Slide forward the DSP Assembly.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE
Set a torque 30.4 in-lb for box torque wrench when fasten the 5/8 inch nuts and the washers
(item 3).
Set a torque to 9 in-lb for torque driver when fasten the TORX T10 screws (item 4).
70
torque screwdriver, T15, and T20
torque screwdriver, T10 (set to 9 in-lb)
box torque wrench, 5/8 inch (set to 30.4 in-lb)