A9 Lf Source Bias Module Replacement; Tools Required; Removal Procedure; Replacement Procedure - Agilent Technologies E5061B Service Manual

Network analyzers
Hide thumbs Also See for E5061B:
Table of Contents

Advertisement

Replacement Procedure

A9 LF Source Bias Module Replacement

A9 LF Source Bias Module Replacement

Tools Required

Removal Procedure

Refer to Figure 4-10 for this procedure.
Step 1. Remove the Outer Cover as described in "Outer Cover Replacement" on page 54.
Step 2. Remove the Front Panel as described in "Front Panel Replacement" on page 56.
Step 3. Disconnect the cable (item 1) from the A9 LF Source Bias Module.
Step 4. Disconnect the three RF cables (item 2) from the A9 LF Source Bias Module.
Step 5. Remove the two TORX T10 screws (item 3) fastening the A9 LF Source Bias Module.
Step 6. Slide forward the A9 LF Source Bias Module.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE
Set a torque to 9 in-lb for torque driver when fasten the TORX T10 screws (item 3).
72
torque screwdriver, T15, and T20
torque screwdriver, T10 (set to 9 in-lb)
flat edge screwdriver
Chapter 4