Replacement Procedure

Handle Assembly Replacement

Handle Assembly Replacement

Tools required

torque screwdriver T20

Removal Procedure

Refer to Figure 4-25 for this procedure.
Step 1. Remove the four TORX T20 screws (item 1) fastening the Handle Assembly.
Step 2. Remove the end caps (both side) (item 2) from the Handle Assembly.
Step 3. Detach parts at both sides (item 3) of the Handle Assembly.
Step 4. Remove the strap cover (item 4) of the Handle Assembly.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.