Agilent Technologies E5061B Service Manual page 47

Network analyzers
Hide thumbs Also See for E5061B:
Table of Contents

Advertisement

Troubleshooting
Performance Test failure Troubleshooting
Table 2-9
Performance Test failure Troubleshooting information (Option
115/117/135/137/215/217/235/237)
1
2
3
4
5
6
7
8
9
10
11
###: Most suspicious assembly
##: Suspicious assembly
#: Possible faulty assembly
46
First failed test
Frequency accuracy test
RF output level accuracy test
RF output level linearity test
Crosstalk & system dynamic range
test
Noise Floor Test
Trace Noise Test
Dynamic Accuracy Test
Compression
IF linearity at 1.195GHz
IF linearity at LF
Uncorrected System Performance
test
Probable faulty board
assembly or parts
A11 Source
A3/A4
Module
Receiver
Module
###
#
###
#
###
#
##
###
##
###
#
###
###
###
###
#
###
Chapter 2