A1 Source Module Replacement; Tools Required; Removal Procedure; Replacement Procedure - Agilent Technologies E5061B Service Manual

Network analyzers
Hide thumbs Also See for E5061B:
Table of Contents

Advertisement

Replacement Procedure

A1 Source Module Replacement

A1 Source Module Replacement

Tools Required

Removal Procedure

Refer to Figure 4-13 for this procedure.
Step 1. Remove the Outer Cover as described in "Outer Cover Replacement" on page 54.
Step 2. Remove the Front Panel as described in "Front Panel Replacement" on page 56.
Step 3. Remove the A2 Receiver Module as described in "A2 Receiver Module Replacement" on
page 74.
Step 4. Disconnect the four RF cable connectors (item 1) from the A1 Source Module and the A9
LF Source Bias Module.
Step 5. Remove the two TORX T10 screws (item 2) fastening the A1 Source Module.
Step 6. Slide forward the A1 Source Module.
Step 7. Remove the four TORX T10 screws (item 3) fastening bracket and angle for the A1 Source
Module.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE
Set a torque to 9 in-lb for torque driver when fasten the TORX T10 screws (item 2, and
item 3).
78
torque screwdriver, T15 and T20
torque screwdriver, T10 (set to 9 in-lb)
open torque wrench, 5/16 inch
flat edge screwdriver
Chapter 4