Post-Repair Procedures
Post-Repair Procedures
Post-Repair Procedures
Table 5-1 Post Repair Procedures lists the required procedures that must be performed
after the replacement of an assembly. These are the recommended minimum procedures to
ensure that the replacement is successfully completed.
Table 5-1
Post-Repair Procedures (Option 3L5)
Replaced
Required Adjustments
Assembly or
Correction Constants (CC)
Part
A1 Source
Perform the following required
Module
adjustments using "A1 Source Module" in
Spot Adjustment of the program.
104
Verification
"To Execute the Diagnostic Test" on page 31
Frequency Accuracy Test
RF Output Level Accuracy and Flatness Test
RF Output Level Linearity Test
Trace Noise Test
Crosstalk & System Dynamic Range Test
Dynamic Accuracy Test
Uncorrected System Performance Test
LF Output Level Accuracy Test (Opt 3L5)
LF Output Level Linearity Test (Opt 3L5)
Input Impedance Test (Opt 3L5)
Absolute Amplitude Accuaracy Test (Opt
3L5)
Magnitude Ratio/Phase Frequency Response
Test (Opt 3L5)
Trace Noise Test (Opt 3L5)
Crosstalk and Noise Level Test (Opt 3L5)
Magnitude Ratio/Phase Dynamic Accuracy
Test (Opt 3L5)