Table 5-2
Post-Repair Procedures
Replaced
Required Adjustments
Assembly or
Correction Constants (CC)
Part
A11 Source
Perform the following required
Module
adjustments using "A11 Source Module"
in Spot Adjustment of the program.
A3/A4 Receiver
Perform the following required
adjustments using "A3/A4 Receiver
Module
Module" in Spot Adjustment of the
program.
A10 Analog
No adjustment needed
Motherboard
A50 CPU
No adjustment needed
module
DSP Card
Perform the following required
adjustments using "DSP board" in Spot
Adjustment of the program.
Writing ID
Front Panel Board
No adjustment needed
Handler I/O
No adjustment needed
Board
Chapter 5
Post-Repair Procedures
Post-Repair Procedures
Verification
"To Execute the Diagnostic Test" on page 31
Frequency Accuracy Test
RF Output Level Accuracy Test
RF Output Level Linearity Test
Crosstalk & System Dynamic Range Test
Noise Floor Test
Trace Noise Test
Dynamic Accuracy Test
Compression Test
IF Linearity at 1.195 GHz Test
IF Linearity at LF Test
Uncorrected System Performance Test.
"To Execute the Diagnostic Test" on page 31
Frequency Accuracy Test
RF Output Level Accuracy Test
RF Output Level Linearity Test
Crosstalk & System Dynamic Range Test
Noise Floor Test
Trace Noise Test
Dynamic Accuracy Test
Compression Test
IF Linearity at 1.195 GHz Test
IF Linearity at LF Test
Uncorrected System Performance Test.
Inspect the Booting Process
When the symptom of failure is out of
specification, perform the performance test.
Inspect the Booting Process
Inspect the Booting Process
Executed all of the performance test.
"To Check the Front Panel" on page 38
"To Check the External Trigger Input" on
page 41