5
First steps
Pos: 32 /TD/Überschriften/5.1 Inbetriebnahme @ 0\mod_1185342823812_7011.doc @ 7099 @
5.1.
Commissioning
Pos: 33 /TD/Erste Schritte/testo 330/Testo 330 Inbetriebnahme @ 6\mod_1278931859503_7011.doc @ 65226 @
The measuring instrument is supplied with a rech. batt. already
fitted.
> Charge the rech. batt. fully before using the measuring
instrument, see Charging the rech. batt. page 24.
Pos: 34 /TD/Überschriften/5.3 Produkt kennenlernen @ 0\mod_1185342901015_7011.doc @ 7100 @
5.2.
Getting to know the product
Pos: 35 /TD/Erste Schritte/testo 330/testo 330 Produkt kennenlernen Netzteil_Akkus_Batt @ 6\mod_1278932047863_7011.doc @ 65258 @
5.2.1.
Mains unit / rech. batt.
If the mains unit is connected, the measuring instrument is
automatically powered from the unit.
5.2.1.1.
Changing the rech. batt.
✓ The measuring instrument must not be connected to a mains
socket via the mains unit. The instrument must be switched off.
Change the rech. batt. within 5 minutes so that device settings (e.g.
date / time) are not lost.
1. Place the measuring instrument on its front.
2. Remove the service cover: Take hold of it at the markings
(arrows) using the index finger and thumb, press slightly, fold up
and remove.
3. Open the battery lock: Press the gray key and push in direction
of arrow.
4. Remove the battery and insert a new rech. batt. Only use the
Testo rech. batt. 0515 0107!
5. Close the battery lock: Press the gray key and push against
direction of arrow until the rech. batt. engages.
5 First steps
23