Cleaning; Dts1 Dielectric Test Sample Box - Omicron Lab SPECTANO 100 User Manual

Hide thumbs Also See for SPECTANO 100:
Table of Contents

Advertisement

6.1.2

Cleaning

To clean SPECTANO 100 and its accessories, use a cloth dampened with
isopropanol alcohol.
6.2

DTS1 dielectric test sample box

DANGER
Death or severe injury caused by high voltage or current
Do not clean SPECTANO 100 when the device is powered.
Do not clean SPECTANO 100 when connected to the test object.
Before cleaning SPECTANO 100 and its accessories, always switch the
device off and disconnect the test object, accessories and connection
cables.
DANGER
Death or severe injury caused by high voltage or current
Use the DTS1 box only together with the SPECTANO 100 test device and
the delivered accessories and cables.
Read the safety instructions in 2 "Safety instructions" in this User Manual
before using the DTS1 box together with the SPECTANO 100 test device.
TS1 box together with the
NOTICE
Equipment damage possible
damage possible
Do not supply a continuous input current > 50 mA
supply a conti
SPECTANO 100 input can causes damages at the device.
ANO 100 inp
0
Always configure
Always configure the max. test output voltage and the input test frequency
for the conne
for the connected test object so that the maximum input current will not be
reached.
eac
For information on how to configure the output signals for testing with the
For inf
DTS1 dielectric test sample box, see "Technical data" in the
DTS
S
SPECTANO 100 User Manual.
e a cloth dampened with
a cloth dampened with
box
high voltage
ether with th
PE
and cables.
ons in 2 "Safe y instr
Device description
switch the
connection
. An overload at the
peak
19