To test the CAL OUTPUT ports (oscilloscope)
Testing the CAL OUTPUT ports does not check a specification, but does check the
following:
• DC CAL OUTPUT voltage
• AC CAL OUTPUT voltage
This test verifies that the CAL OUTPUT voltages are operating within limits, so that
they can provide accurate calibration for the instrument operational accuracy
calibration and probe calibration.
Equipment Required
Equipment
Digital Multimeter
Cable
Adapter
Critical Specifications
0.1 mV resolution, better than 0.005% accuracy
BNC (m)(m) 48-inch
BNC
(II
to Dual Banana Plug
Recommended
ModelJPart
HP 3458A
HP 8120-1840
HP 1251-2277
Set up the equipment
Turn on the equipment required and the logic analyzer. Let them warm up for
30 minutes
if
you have not already done so.
3-65