CURSOR MEASUREMENT

The cursor can be used to measure the X axis (time) and the Y axis (voltage) of the
selected waveform. Press the CURSOR button to enter the cursor measurement
menu.
Time Measurement
Press the CURSOR button to enter the cursor measurement menu. Then press the
F1 key to select the type to measure.
Press F2 to select the source of the time measurement channel and set the pattern
to independent.
The multi-purpose control can adjust the vertical cursor AX, press it to switch to
cursor BX.
The measured values are shown in the upper left corner of the waveform display
area. The value of BX-AX is the time measurement, and 1/|BX-AX| is the reciprocal
of time, or frequency.
For a periodic signal, if AX and BX are set at the rising edge of adjacent cycles,
then BX-AX is the signal's period, and 1/|BX-AX| is the frequency.
Parameters can also display the voltage value at cursor's current position. That is
AY, BY, and BY-AY.
The multi-purpose control can be used to adjust the BX and AX positions when
setting is set to trace using the F3 key.

VOLTAGE MEASUREMENT

The method for voltage measurement is similar to the method for time
measurement, only that the vertical cursor becomes the horizontal cursor.
Press the CURSOR button to enter the cursor measurement menu. Then press the
F1 key to select the type to voltage.
Press F2 to select the source of the voltage measurement channel and set the
pattern to independent.
The multi-purpose control can be used to adjust the horizontal cursor AY on the
screen, press the multi-purpose control to switch to cursor BY. The value of BY-AY
is the voltage measurement (V).