Press F1 to select GROUND.
Both AC and DC quantities of the signal being measured are now intercepted.
The waveform is not displayed in this mode but the signal remains connected to the
channel circuit.
Channel bandwidth setup
If for example a signal is applied to CH1 which is a pulse signal that contains high
frequency oscillation.
Press CH1 to select Channel1.
Press F2 to set the BANDWIDTH LIMIT OFF so it is set up as full bandwidth.
The signal being measured can now pass through even if it contains high frequency
quantities.
Press F2 to set BANDWIDTH LIMIT ON so that frequency quantities higher than
20MHz in the signal being measured will be limited.
Vertical Volts/Div adjustment setup
The VOLTS/DIV range of the vertical deflection factor can be adjusted either in
coarse or fine tune mode.
In COARSE TUNE the VOLTS/DIV range is 1mV/div~2V/div. Tuning is in steps of
1-2-5. For example: 10mV - >20mV - >50mV - >100mV.
In FINE TUNE mode it adjusts in 1% of the current amplitude value steps.
For example: 10.00mV—>10.10mV—>10.20mV—>10.30mV.
Probe rate setup
To match the probe attenuation factor setup, it is necessary to set up the probe
attenuation factor in the channel operation menu accordingly.
For example when the probe attenuation factor is 10:1, set the probe attenuation
factor at 10X in the menu. This principle applies to other values to ensure the
voltage reading is correct.
The coefficients can be set to 0.001X, 0.01X, 0.1X, 1X, 10X, 100X, and 1000X.
Waveform inversion (Reverse Phase)
The displayed signal is inverted 180 degrees with respect to the ground level.
The vertical identification will display a reverse phase logo ↓.