Digital phosphor storage oscilloscope
Table of Contents

Advertisement

Functions
Options
Type
Format
XY 1&2
XY 3&4
100ms /
Duration
500ms / 1s
time
/ 2s /5s /
Continuous
Grid
1%~100%
Brightness
Waveform
1%~100%
Brightness
XY Pattern
The XY mode display is also called Lissajous graph.
When XY 1&2 are selected, the CH1 signal will be entered on the horizontal axis
(X), and the CH2 on the vertical axis (Y).
When XY 3&4 are selected, the CH3 signal will be entered on the horizontal axis
(X), and the CH4 on the vertical axis (Y). UPO2XX4CS only.
In X-Y mode, when CH1 or CH3 are activated, use the horizontal POSITION control
to adjust the graph horizontally.
When CH2 or CH4 are activated, use the horizontal POSITION control to adjust the
graph vertically.
The vertical SCALE control to be used to adjust each channel's amplitude level.
The horizontal SCALE control can be used to adjust the time position in order to
obtain a well displayed Lissajous graph.
Vector
Point
YT
Display Lissajous graph of CH3~CH4 waveforms (UPO2XX4CS
Close
200ms
10s
Set the grid brightness, Use multi-purpose control to adjust
Set the waveform brightness, Use multi-purpose control to adjust
Description
Using lines to display between samples
Displays points directly
Display level on time scale
Display Lissajous graph of CH1~CH2 waveforms
Displayed waveform refreshes at normal rate
Displayed waveform refreshes after specified time
Waveform refreshes at a continuous rate
31
only)
settings
settings