ÄKTA go Operating Instructions 29360951 AC
Action
Connect a 25 to 30 mL syringe to the purge valve of one of the pump heads.
Open the purge valve by turning it counter-clockwise one and a half turns.
Draw liquid slowly into the syringe until liquid reaches the pump and the inlet
tubing is filled with liquid.
Close the purge valve by turning it clockwise. Disconnect the syringe and
discard its contents.
Repeat steps 3 to 6 for each inlet that is to be used during the run.
5 Prepare the system for a run
5.2 Prime inlets and purge pump heads
65