7 | Warning and error messages
7 Warning and error messages
Warnings
Error messages
Touchscreen
7.1 List of warning and error messages
Type
Notification
W102
Timeout during
communication with
EEPROM in internal IO
module
W104
One EEPROM parameter
has been initialized
52 / 74
During operation, the display shows information that helps you operate the device.
Measurement values are displayed along with current device modes, operating
instructions as well as warnings and error messages. The device is equipped with
extensive self-diagnostic functions. If the electronics detect a faulty state, the device
will show this as far as possible on the display and will interrupt operation.
Warnings warn of device states that can impair the accuracy of measurements. To
perform measurements, confirm active warnings with the "Clear" button.
Errors are events that force the interruption of the operation. The error message
consists of a number and a descriptive text. Once you have rectified the cause of the
error, continue operation by pressing the button "Clear".
You will find an overview of possible errors and warnings on the touchscreen:
►
> Help > Errors and warnings
Possible sources of
error
The EEPROM in the
internal IO module is
defective or not present
A new parameter was
introduced by a
software update
The EEPROM in the
internal IO module is
defective
ELT3000-Battery Leak-Detector-Operating-instructions-mina95en1-10-(2007)
Remedy
• Contact customer service
• Confirm the warning message
• Check that the message does not appear
when you restart the device
• Check whether the factory setting of the
new parameter corresponds to your
application
• Confirm the warning message
• Check if the message occurs each time
when you restart the device
• Contact customer service
INFICON
Need help?
Do you have a question about the ELT3000 and is the answer not in the manual?