Inficon ELT3000 Translation Of The Original Operating Instructions page 61

Battery leak detector
Table of Contents

Advertisement

INFICON
Type
Notification
W600
Calibration factor too low
W601
Calibration factor too high
W605
Signal of calibration leak too
low
W630
Calibration request
W660
Calibration - Offset too high
E661
Calibration - Signal too low
or offset too high
E709
Temperature of basic unit
too low
W710
Temperature of basic unit
too high
E711
Maximum temperature of
basic unit exceeded
ELT3000-Battery Leak-Detector-Operating-instructions-mina95en1-10-(2007)
Possible sources of
error
Incorrect value entered
at calibration
Incorrect specimen
inserted
ZERO measurement
error
Incorrect value entered
at calibration
Incorrect specimen
inserted
ZERO measurement
error
Incorrect value entered
at calibration
Incorrect specimen
inserted
ZERO measurement
error
Operation mode or
mass has changed
Calibration leak during
zero measurement in
chamber
Background too high
Calibration leak during
zero measurement in
chamber
Signal of calibration
leak too small
Ambient temperature is
too low
The ambient
temperature is too high
The ambient
temperature is too high
Warning and error messages | 7
Remedy
• Repeat the calibration
• Repeat the calibration
• Repeat the calibration
• Repeat the calibration
• Repeat the calibration
• Repeat the calibration
• Repeat the calibration
• Repeat the calibration
• Repeat the calibration
• Perform a calibration
• Repeat the calibration
• Use the purge function to reduce the
background
• Repeat the calibration
• Use a different calibration leak
• Increase the temperature in the
environment where the device is located
• Reduce the temperature in the
environment where the device is located
• Reduce the temperature in the
environment where the device is located
61 / 74