Inficon ELT3000 Translation Of The Original Operating Instructions page 53

Battery leak detector
Table of Contents

Advertisement

INFICON
Type
Notification
W106
Several EEPROM
parameters have been
initialized
E107
Internal IIC communication
error
W110
Real-time clock was reset!
Enter date and time
W122
No response from bus
module
ELT3000-Battery Leak-Detector-Operating-instructions-mina95en1-10-(2007)
Possible sources of
error
A software update
introduced new
parameters
The EEPROM has
been replaced in the IO
module
The EEPROM in the
internal IO module is
defective
Internal IIC
communication error
The real-time clock has
not been set
Battery is discharged or
defective in internal IO
module
Real-time clock
defective
Connection to BUS
module interrupted
Bus module defective
Bus module connection
on the device defective
Warning and error messages | 7
Remedy
• Confirm the warning message
• Check that the message does not appear
when you restart the device
• Check whether the factory setting of the
new parameters corresponds to your
application
• Confirm the warning message
• Check that the message does not appear
when you restart the device
• Check whether the factory setting of the
new parameters corresponds to your
application
• Confirm the warning message
• Check if the message occurs each time
when you restart the device
• Contact customer service
• Contact customer service
• Enter the correct date and time
• Check that the message does not appear
when you restart the device
• Contact customer service
• Contact customer service
• Check the connection to the bus module
• Replace the connection cable to the bus
module
• Replace the bus module
• Contact customer service
53 / 74