Cleaning And Maintenance; Gas Control Unit: Cleaning The Housing; Gas Control Unit: Replace Hoses - Inficon ELT3000 Translation Of The Original Operating Instructions

Battery leak detector
Table of Contents

Advertisement

INFICON

8 Cleaning and maintenance

8.1 Gas Control Unit: Cleaning the housing

8.2 Gas Control Unit: Replace hoses

ELT3000-Battery Leak-Detector-Operating-instructions-mina95en1-10-(2007)
All cleaning and maintenance work described here must be carried out without
opening the device cover!
DANGER
Risk of death from electric shock
There are high voltages inside the device. Touching parts where electrical voltage is
present can result in death.
► Disconnect the device from the power supply prior to any cleaning and maintenance
work. Ensure that the electrical supply cannot be switched back on unintentionally.
► Do not open the device covers!
The housing for the device consists of a painted metal housing and an optional
measuring chamber made of aluminum.
1
Only use water for moistening.
2
Avoid cleaning agents that contain alcohol, fat or oil.
3
Make sure that the device is disconnected from the power supply by
disconnecting the power supply plug.
4
Wipe the housing with a soft damp cloth.
5
When cleaning the measuring chamber, use an agent that is suitable for
aluminum surfaces (for example, a gentle household cleaner). Do not use
solvents that can attack the painted metal housing.
DANGER
Electrolyte or battery acid is corrosive.
► Avoid contact with skin, eyes, or clothing.
► Wear suitable protective clothing, especially gloves, apron and face protection.
► Observe the information in the respective MSDS.
► Rinse off electrolyte or acid splashes immediately with clean water.
► Consult a physician if necessary.
Cleaning and maintenance | 8
63 / 74

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ELT3000 and is the answer not in the manual?

Table of Contents