JUMO flowTRANS MAG S Series Safety Manual page 45

Electromagnetic flowmeter for the process industry and hygienic applications
Table of Contents

Advertisement

5 Ex-relevant data for operation in zones 2, 21 and 22
Table 3: Medium temperature (Ex data) for model 406012/2-1, 406013/2-1, 406015/2-1, 406016/2-1
IMPORTANT (NOTE)!
If the installation location for the device is classified as a potentially explosive area for gases and dust,
the temperature data in the "Gas & dust" columns in the table must be taken into consideration.
If the installation location for the device is only classified as a potentially explosive area for gases, the
temperature data in the "Gas" column in the table must be taken into consideration.
IMPORTANT (NOTE)!
Not thermally insulated: the sensor is not surrounded by pipe insulating material.
Thermally insulated: the sensor is surrounded by pipe insulating material.
Ambient temperature
c
(-40 °C)
-20 °C to +40 °C
Not thermally
insulated
Gas
Gas
&
Dust
T1 130 °C 130 °C -
T2 130 °C 130 °C -
T3 130 °C 130 °C -
T4 130 °C 130 °C -
T5 95 °C
95 °C
T6 80 °C
80 °C
a
Applies for model 406012/2-1, 406013/2-1
b
Applies for model 406015/2-1, 406016/2-1
c
Low temperature version (optional)
c
(-40 °C)
Thermally
Not thermally
insulated
insulated
Gas Gas
Gas
Gas
&
&
Dust
Dust
-
130 °C 130 °C -
-
130 °C 130 °C -
-
130 °C 130 °C -
-
130 °C 130 °C -
-
-
95 °C
95 °C
-
-
80 °C
80 °C
-20 °C to +50 °C
(-40 °C)
Thermally
Not thermally
insulated
insulated
Gas Gas
Gas
&
Dust
-
110 °C
120 °C
-
110 °C
120 °C
-
110 °C
120 °C
-
110 °C
120 °C
-
-
95 °C
-
-
80 °C
c
-20 °C to +60 °C
Thermally
insulated
Gas
Gas Gas
&
&
Dust
Dust
a
110 °C -
-
b
a
110 °C -
-
b
a
110 °C -
-
b
a
110 °C -
-
b
95 °C
-
-
80 °C
-
-
45