4 Ex relevant data for operation in zones 1, 21 and 22
4.2

Electrical data for operation in zones 1, 21, 22

4.2.1

Devices with HART protocol

When operating in potentially explosive areas, observe the following electrical data for the transmitter
signal inputs and outputs. For the current output version (active/passive), see the identification marking
in the device's terminal box.
Model: 406012/1-1, 406013/1-1, 406015/1-1, 406016/1-1, 406018/2-1 (dual-compartment housing),
406019/2-1 (dual-compartment housing)
Inputs and U
Outputs
Current
output
Active
Terminal
31/32
Current
output
passive
Terminal
31/32
Digital
output
DO2 pas-
sive
Terminal
41/42
Digital
output
DO1 pas-
sive
Terminal
51/52
Digital
input
d
DI
passive
Terminal
81/82
a
Intrinsically safe single-channel or multichannel barriers (supply isolators) with resistance characteris-
tic curve must be used.
b
For "active" current output.
c
For "passive" current output.
d
Only available in conjunction with passive current output.
All inputs and outputs are galvanically isolated from each other and from the voltage supply.
Operating
Ex-data – ignition protection type Ex i, IS
data
I
U
U
N
N
O
[V]
[mA]
[V] [V] [mA] [mA] [mW] [mW] [nF] [nF] [nF]
30
30
20 60 100
30
30
60
30
220
60
30
220
60
30
10
60
a
I
I
P
I
O
I
O
425
500
500
b
425
c
500
b
425
c
500
500
a
P
C
C
C
I
O
I
OPA
2000 210 8.4
195
2000
8.4
2000
3.6
2000
3.6
2000
3.6
C
L
L
IPA
O
I
[nF] [mH] [mH]
24
6
0.065
24
170
3.6
170
3.6
170
3.6
170
29