F.3.8
Fault opto film
This alarm message is always related to the optoswitch S16 that
carriage, as the alarm Cassette (see the description in this sec-
tion of the Service Manual) but it indicates that the optoswitch
is broken.
Press the RESET pushbutton on the control panel to cancel the
alarm.
F.3.9
Collimator
The position of the collimator is detected by three optoswitches
(S21,S22 and S30) as listed in the table below:
S21
S22
S30
ON
ON
ON
ON
ON
OFF
ON
OFF
ON
ON
OFF
OFF
OFF
ON
ON
OFF
ON
OFF
OFF
OFF
ON
OFF
OFF
OFF
This alarm is generated when the collimator is in an intermediate
position, i.e. between transcan and panoramic, while it is moved
to change its position according to the projection desired (pan-
oramic, transcan, teleradiography).
When the correct position is reached, the alarm is automatically
reset.
The alarm is also generated if the collimator is in the transcan
position but the transcan option is not present in the system.
If the alarm stays on it may mean:
1.
one of the collimator optoswitches S21, S22, S30 is damaged
(see principle diagram Z1-11)
2.
damaged (see principle diagram Z1-11)
3.
If the converter board PQ1 needs to be replaced,
the service programs Tube calibration and Vf tube
editing must be performed (see the relevant para-
graphs of this section for more information).
Gendex Dental Systems 4519-190-71193-7
MESSAGE /CONDITION
Optoswitches collimator fail
Optoswitches collimator fail
Optoswitches collimator fail
Collimator in Ceph position
Collimator in Pan position
Collimator in Tscan position
Optoswitches collimator fail
Collimator
F-5
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
ORTHORALIX 9200 Service Manual
F.3.10
Tubehead
The position of the tubehead is detected by two optoswitches
(S19 for the teleradiographic position and S20 for the panoramic/
transcan position) as listed in the table below:
S19
S20
CONDITION
Tubehead in Pan/
OFF
ON
tscan position
Tubehead in Ceph
ON
OFF
position
ON
ON
Error condition
OFF
OFF
Error condition
This alarm is generated when the position of the tubehead is not
congruent with the position of the collimator.
When the correct position is reached, the alarm is automatically
reset.
If the alarm is always present:
1.
one of the tubehead optoswitches S19 or S20 is damaged
(see principle diagram Z1-11)
2.
wiring between the optoswitches S19 - S20 and the converter
board damaged (see principle diagram Z1-11)
3.
If the converter board PQ1 needs to be replaced,
the service programs Tube calibration and Vf tube
editing must be performed (see the relevant para-
graphs of this section for more information).
MESSAGE
None
None
Optoswitches
tubehead fail
Tubehead
(03.0)E