9.
During the mounting of the new tubehead, please take note
of the following:
nections on the terminal block 600
on the terminal block 600, otherwise there is the possibility
to unlock the internal connections.
10.
After the replacement of the tubehead, the following adjust-
ments are necessary:
version - see Section C and F).
phantom (see Section C and F).
Gendex Dental Systems 4519-190-71193-7
-
F-45
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
ORTHORALIX 9200 Service Manual
F.5.7
Replacing the PC1 control board
A) For Orthoralix 9200/Plus/Ceph
To replace the PC1 control board, proceed as follows:
1.
remove the mirror panel (a) by unscrewing the two Allen
screws (b).
2.
disconnect the connector X31 from the PC1 board.
Pay attention to do not damage the connectors and the
wiring.
3.
disconnect all the others connectors from the PC1 board.
4.
remove the PC1 board (c) by unscrewing the five screws
(d).
(03.0)E