F.5.2
Replacing converter cable
For the replacement of the converter cable follow the procedure
below:
Before starting the replacement of the converter
cable, be sure that no dangerous voltages are
present.
1.
Remove the top longer cover (a) of the rotating arm by
unscrewing the six screws and the top short cover (b) by
unscrewing the four screws.
2.
Remove the top cover (e) of the Z-carriage by unscrewing
the six Allen screws.
3.
Remove the lower cover (c) of the Z-carriage by unscrewing
the four screws.
4.
Remove the central cover (d).
Gendex Dental Systems 4519-190-71193-7
F-35
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
ORTHORALIX 9200 Service Manual
5.
Remove the metallic protection (f) of the PN1 board by
6.
Remove the headrest cover (h) by unscrewing the four screws
(i).
7.
Disconnect the connector X13 from the headrest.
8.
Remove the headrest assembly (l) by unscrewing the three
Allen screws (m).
(06.0)E