9.
Remove the plastic disk (n) by unscrewing the six screws (o)
and then remove the three plastic rings (p).
Before disconnecting all the connectors of the multipole
cable, sign two reference marks on the old multipole cable:
1. one at the entrance of the cable into the hole for the rotation
2. one at the exit from the hole
This to ensure the right cable length inside the rotation part
(see figure in the next page).
10.
Disconnect all the connectors from the PC1 board and from
the PA7 board.
NOTE
pay attention not to damage the connectors and the wiring
11.
Remove the PC1 board (q) by unscrewing the five screws (r);
remove the PA7 board from the PC1 board by unscrewing
the three screws
12.
Dismount the support plate (s) by unscrewings the five
screws (t).
13.
Gendex Dental Systems 4519-190-71193-7
F-34
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
14.
Loosen the two screws (u) of the cable retainer (v).
15.
Remove all the necessary cable retainers of the multipole
cable; disconnect all the necessary connectors and finally
remove the multipole cable itself.
16.
Take the new multipole cable, lay it down near the old one
and sign on it the same reference marks placed on the old
one.
17.
Mount the new multipole cable and fix it in the same fixation
points.
18.
Proceed backwards for remounting.
After the mounting of the headrest, verify the
ment of patient headrest (pan) in this Section of
the Service Manual.
Adjust-
Entrance