F.6.5
Adjusting patient headrest (PAN)
Using a plumb line, verify that the middle position of the headrest
assembly is aligned with the middle position of the bite block (see
reference middle line on the bite block itself and the figure below).
If the headrest assembly is not aligned with the bite block, pro-
ceed as follows:
1.
Remove the cover (a) of the headrest by unscrewing the four
screws (b).
2.
Unlock the three fixation Allen screws (c).
3.
Adjust the headrest assembly in the proper position and
tight the three fixation Allen screws.
4.
Mount back the cover (a).
Reference middle line
Gendex Dental Systems 4519-190-71193-7
F-65
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
ORTHORALIX 9200 Service Manual
(03.0)E