C.5.6
Aec ceph calibration
The parameter used for the calibration of the ceph AEC sensor and
cassette/screens is written on a label should be stuck on the ceph
where the following information are present:
S/N
is the S/N of the PA5 board
Cassette
is the cassette and screens used for the calibra-
C
is the calibration parameter for the AEC
Date
date of the calibration
The ceph AEC sensor and cassette are already calibrated in fac-
Manual for more details.
Please record this calibration parameter also in the Configura-
tion Chart.
C.5.7
Configuration chart
C.5.8
AEC Calibration with Denoptix
When the Denoptix system is used in combination with an Ortho-
Gendex Dental Systems 4519-190-71193-7
C-10
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
C.6
Functional test
C.6.1
Performance test
1.
Switch on the system with the mains ON/OFF switch.
2.
3.
Check all the operator functions with the help of the User
Manual.
4.
ditions for panoramic exposure and check that the green
"READY" lamp is on.
5.
for transcan exposure and check that the green "READY"
lamp is on.
6.
7.
and check that the green "READY" lamp is on.
The following tests require X-ray exposure. Take
all the necessary radiation precautions.
8.
sure by means of the handswitch. Check that the X-ray buzzer
indication is generated.
9.
Make a new exposure and release the X-ray pushbutton
before the end of the exposure: check that the X-rays and
by the "dead-man" function.
10.
and "TUBEHEAD" operator alarms described in section F.
C.6.2
Customer test
standards perform the tests described in section D: Acceptance.
-
-
-
-