F.8
Converter inputs read
A) Orthoralix 9200
On the display the following message will appear:
CONV INPUTS READ
XY 0 0 21
where the numbers X and Y may have the following values,
converter board:
X = 1
when the tubehead is in the panoramic position
X = 2
when the tubehead is in the ceph position
X = 3
when rotating the tubehead from panoramic to ceph
position or viceversa
Y = 8
when the collimator is in the transcan position and the
panoramic cassette is inserted in the carriage
Y = 9
when the collimator is in the transcan position and the
panoramic cassette is not inserted in the carriage
Y = A
when the collimator is in the panoramic position and
the panoramic cassette is inserted in the carriage
Y = B
when the collimator is in the panoramic position and
the panoramic cassette is not inserted in the car-
riage
Y = 4
when the collimator is in the ceph position and the
panoramic cassette is inserted in the carriage
Y = 5
when the collimator is in the ceph position and the
panoramic cassette is not inserted in the carriage
Y = 0
when moving the collimator from one position to an-
other (i.e. from panoramic to ceph) and the panoramic
cassette is inserted in the carriage
Y = 1
when moving the collimator from one position to an-
other (i.e. from panoramic to ceph) and the panoramic
cassette is not inserted in the carriage
Gendex Dental Systems 4519-190-71193-7
F-73
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
ORTHORALIX 9200 Service Manual
B) Orthoralix 9200 DDE
On the display the following message will appear:
CONV INPUTS READ
XY 0 0 22
where the numbers X and Y may have the following values,
converter board:
X = 0
when in ERROR
X = 1
when the tubehead is in the old ceph position (error
condiction)
X = 2
when the tubehead is in the panoramic position
X = 3
when rotating the tubehead from panoramic to ceph
position or viceversa
Y = 0 or 1 when moving the collimator from one position to
another
Y = 4 or 5 when the collimator is in the ceph position
Y = 8 or 9 when the collimator is in the Transcan position
Y = A or B when the collimator is in the panoramic position