ORTHORALIX 9200 Service Manual
13.
four Allen screws (a) in the ceph arm and move the ceph
secondary collimator assembly (b) in the desired direction.
14.
If the X-ray field shall be vertically adjusted, remove first the
rear cover (a) and then the front cover of the ceph secondary
collimator, unlock the four screws (b) and move the lead plate
(c) in the vertical desired position.
Gendex Dental Systems 4519-190-71193-7
F-58
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
ary collimator, the alarm Ceph collimator appears
on the display, see Step (2) point (14).
-
(04.0)E